Design and development for amelioration of primary water flow standard and calibration systems

  • Shiv Kumar Jaiswal(corresponding author)
    ,
  • Chatar Singh
    ,
  • Sanjeet Kumar
    ,
  • Anshul Varshney
    ,
  • ,
  • Sanjay Yadav
*Corresponding author for this work
Research output:
Contribution to journal
Article
Research
Peer-review

About the research

<p>The present study outlines the efforts made to improve national primary water flow standards and calibration systems through design and development. The facility has been designed and developed in accordance with ISO 4185 standard in the flow range 0.03 m<sup>3</sup>/h to 650 m<sup>3</sup>/h to calibrate various types of flow meters up to DN200 (Nominal diameter) using 12 kg, 300 kg, 3000 kg, and 6000 kg weighing systems. In the flow range up to 530 m<sup>3</sup>/h, the expanded uncertainty in flow meter calibration in totalized mode is found to be ±0.01% to ±0.025% (k = 2), whereas it is ±(0.03–0.05) % (k = 2) up to DN200 size (test rigs) for mass flow rate (MFR) and volume flow rate (VFR) in the flow range 0.1 m<sup>3</sup>/h to 650 m<sup>3</sup>/h. The measurement uncertainty achieved is comparable to that of state-of-the-art water flow measurement capabilities available at numerous National Metrology Institutes (NMIs). Thus, the present designed and developed system at CSIR-National Physical Laboratory (CSIR-NPL) is a solution to maintain traceability to the users and industries.</p>

Research output information

Output type

Research output:
Contribution to journal
Article
Research
Peer-review

Original language

English

Article number

102201

Journal (Volume, Issue Number)

Flow Measurement and Instrumentation (Volume 86)

Publication milestones

  • Accepted/In press - 09/06/2022
  • Published - 22/06/2022